Dräger Tube -Mercury Vapour 0.1/b

DrägerSKU: CH23101

Price:
Sale price£78.41

Description

Dräger Tubes For Short Term Measurements: Mercury Vapor 0.1/b

Dräger Safety offer a massive range of short-term gas detection tubes for the measurement of various gases. This particular set of tubes is for (Mercury Vapor 0.1/b).

The tubes come in a pack of 10 tubes and have a shelf life of 2 years.

Application Range:

Standard Measuring Range: 0.05 to 2 mg/m3
Number of Strokes n: 40 to 1
Time for Measurement: max. 10 min
Standard Deviation: ± 30 %
Colour Change: pale yellow grey to pale orange

Ambient Operating Conditions:

Temperature: 0 to 40 °C
Absolute Humidity < 20 mg H2O / L

Reaction Principle:

Hg + Cul ~> Cu-Hg-complex

Cross Sensitivity:

Free halogens cause substantial minus errors. It is impossible to measure mercury vapor in the presence of halogens. Arsine, phosphine, hydrogen sulfide, ammonia, nitrogen dioxide, sulfur dioxide and hydrazine in the TLV range do not interfere..

For more information on Dräger-Tubes click the image below.

To view our range of Dräger pumps please click below.

You may also like

Recently viewed