Midas® S2 Sensor Cartridges Arsine (AsH3)5-200 ppb, Germane (GeH4) 70-800 ppb P/N: MIDAS S2-E-ASH

HoneywellSKU: MIDAS S2-E-ASH

Price:
Sale price£321.00

Description

Midas® S2 Sensor Cartridges Arsine (AsH3)5-200 ppb, Germane (GeH4) 70-800 ppb P/N: MIDAS S2-E-ASH

Selectable gases Arsine (AsH3)
 Germane (GeH4)
Cartridge Part Number MIDAS S2-E-ASH 2-year extended warranty
Measuring Range AsH3 0 – 200ppb
GeH4 0 – 800ppb
Default Alarm 1 AsH3 5ppb
GeH4 100ppb
Default Alarm 2 AsH3 10ppb
GeH4 200ppb
LDL, LAL AsH3 5ppb
GeH4 70ppb
Resolution AsH3 1ppb
GeH4 5ppb
Accuracy ≤ ± 5% of measured value
Response Time t62.5 Typical 10 seconds
Sensor Cartridge Life Expectancy 
(Expiration Period)
24 months under typical application conditions
Extendable for 1 year through calibration after 
24 months
Operating Temperature
Effect of Temperature Sensitivity
0°C to +40°C (32°F to 104°F)
≤ ± 25% of measured value at 20°C
Operating Humidity 10 to 90% non-condensing
Operating Pressure 90 – 110kPa
Calibration Gas AsH3 100ppb
GeH4 400ppb
Warm Up Time < 20 minutes
Storage Temperature +5°C to +25°C (+41°F to +77°F)
ISS is not responsible for the data contained within this table.

 

 

You may also like

Recently viewed