Midas® S2 Sensor Cartridges Difluoromethane (CH2F2) ** P/N: MIDAS S2-E-XCF

HoneywellSKU: MIDAS S2-E-XCF

Price:
Sale price£321.00

Description

Midas® S2 Sensor Cartridges Difluoromethane (CH2F2) ** P/N: MIDAS S2-E-XCF

Selectable gases Hexafluorobutadiene (C4F6) 
 Octofluorocyclopentene (C5F8)
 Difluoromethane (CH2F2) (R32)
 R134a (C2H2F4)
 Hexafluoroisobutylene (C4H2F6)
 Hydrogen Fluoride (HF) 
Cartridge Part Number MIDAS S2-E-XCF 2-year extended warranty
Sensor Technology Electrochemical sensor
Measuring Range C4F6 0 - 40ppm
C5F8 0 - 40ppm
CH2F2 0 - 120ppm
C2H2F4 0 - 1000ppm
C4H2F6 0 - 40ppm
HF 0 - 12ppm
Default Alarm 1 C4F6 5ppm
C5F8 5ppm
CH2F2 15ppm
C2H2F4 250ppm
C4H2F6 12ppm
HF 1.5ppm
Default Alarm 2 C4F6 10ppm
C5F8 10ppm
CH2F2 30ppm
C2H2F4 500ppm
C4H2F6 24ppm
HF 3ppm
LDL, LAL C4F6 1.7ppm
C5F8 2ppm
CH2F2 6ppm
C2H2F4 40ppm
C4H2F6 11ppm
HF 1.05ppm
Resolution C4F6 0.2ppm
C5F8 0.2ppm
CH2F2 1ppm
C2H2F4 2ppm
C4H2F6 0.2ppm
HF 0.05ppm
Accuracy ≤ ± 10% of measured value
Response Time t62.5 Typical 60 seconds
Sensor Cartridge Life Expectancy
(Expiration Period)
24 months under typical application conditions
Extendable for 1 year through calibration after
24 months
Operating Temperature
Effect of Temperature Sensitivity
0°C to +40°C (32°F to 104°F)
≤ ± 20% of measured value at 20°C
Operating Humidity 20 to 75% non-condensing
Operating Pressure 90 – 110kPa
Calibration Gas HF 6ppm
Warm Up Time < 20 minutes
Storage Temperature +5°C to +25°C (+41°F to +77°F)